The Medical House PLC
image

VISC-ASI™ Study

Increasingly drug products are being offered as “sustained release” formulations, which provide the benefit of less frequent injections for patients. However these sustained release formulations are often viscous – requiring greater force to achieve an acceptable injection duration, using a narrow gauge needle. This need to apply greater force can result in difficulties for patients who may face challenges associated with dexterity, strength or motor function. Although the assistance of a disposable autoinjector can help overcome these challenges, the use of high forces can result in breakage of glass syringes, with the consequent risk of injury to patients. The VISC-ASI™ disposable autoinjector is specifically designed to enable automated self-injection of viscous drugs. The proprietary Syringe Protection System of the VISC-ASI™ protects glass syringes against the risk of damage from the application of high forces.

Data generated by Bespak Injectables in a study using VISC-ASI™ to deliver a range of liquid viscosities through different needles is summarised below.

To discuss your specific viscous product delivery requirements please click here to contact Bespak Injectables.

Viscosity mPas (Cps)

Needle Gauge

Needle Length (mm)

Volume (ml)

Delivery Time (sec)

35

26

12.5

1

7

35

27

12.5

1

10

85

25

12.5

1

7

110

25

12.5

1

10

220

25

12.5

1

13

400

23

25

1

13